Uni Bremen Home MARUM - Sediment Geochemistry - Leobener Str - D-28359 Bremen - Germany
Marine Geochemistry - Laboratory Methods

inlab methods main

onboard methods main


Access element methods

Sr - Strontium

sample type

marine porewater
digested sediment 1g/L
groundwater
typical concentration
60 µmol/L
5 ppm
2 µmol/L
150 ppb

0.5 µmol/L
50 ppb

Method1 ICP-OES
wavelength 407.771 nm -> wavelength table
method used
yes
yes
yes
dilution
1:10
sample size
>1 mL
> 2mL 
> 2 mL 
detction limit
 50 ppb
5 ppb 
5 ppb  
standard solutions
single element standard 
single element standard, M6
single element standard 
Method 2 ICP-MS
Mass 88 -> Mass Table
method used
yes
yes
yes
dilution
1:100
1:10
1:10
sample size
>1mL
> 1 mL
> 1mL
detction limit
100 ppb
10 ppb
10 ppb
standard solutions
Multielement standard in nitric acid 1% M4, M6, Kraft ME
Multielement standard in nitric acid 1% M4, M6, Kraft ME
Multielement standard in nitric acid 1% M4, M6, Kraft ME
tips + tricks  

page created by M.Kölling & T.Feseker