sample
type
|
marine
porewater
|
digested
sediment 1g/L
|
groundwater
|
| typical
concentration |
0.5 µmol/L
50 ppb
|
0.01 µmol/L
1 ppb
|
0.02 µmol/L
2 ppb
|
| Method1 |
ICP-OES
with USN |
| wavelength |
202.030
nm -> wavelength table |
|
method used
|
no
|
yes
|
no
|
|
dilution |
|
-
|
|
|
sample size |
|
>
2 mL
|
|
| detction
limit |
|
10
ppb
|
|
|
standard solutions |
|
Single
element standard in nitric acid 1%
|
|
| Method
2 |
GF-AAS |
| wavelength |
313.3
nm |
|
method used
|
yes
|
yes
|
yes
|
|
dilution |
1:10
|
1:100
|
-
|
|
sample size |
>
100 µL
|
>
100 µL
|
>
100 µL
|
| detction
limit |
5
ppb
|
2
ppb
|
2
ppb
|
|
standard solutions |
Single
element standard in nitric acid 1%
|
Single
element standard in nitric acid 1%
|
Single
element standard in nitric acid 1%
|
| Method
3 |
ICP-MS |
| Mass |
95
-> Mass Table |
|
method used
|
yes
|
yes
|
yes
|
|
dilution |
1:100
|
1:10
|
1:10
|
|
sample size |
>
1 mL
|
>
1 mL
|
>
1mL
|
| detction
limit |
10
ppb
|
1
ppb
|
1
ppb
|
|
standard solutions |
|
|
|
| tips