Mn - Manganese

sample type

marine porewater
digested sediment 1g/L
groundwater
typical concentration
µmol/L
ppb
µmol/L
ppb

µmol/L
ppb

Method1 ICP-OES
wavelength 257.610 nm -> wavelength table
method used
no
yes
no
dilution
 
-
 
sample size
 
> 2 mL
 
detction limit
 
20 ppb
 
standard solutions
 
Single element standard in nitric acid 1%
 
Method 2 GF-AAS
wavelength 279.5 nm
method used
no
no
yes
dilution    
-
sample size    
> 500 µL
detction limit    
1 ppb
standard solutions    
Single element standard in nitric acid 1%
Method 3 ICP-MS
Mass 55 -> Mass Table
method used
yes
yes
yes
dilution
1:100
1:10
1:10
sample size
> 1 mL
> 1 mL
> 1mL
detction limit
10 ppb
1 ppb
1 ppb
standard solutions
Multielement standard in nitric acid 1% M4, M6, Kraft ME
Multielement standard in nitric acid 1% M4, M6, Kraft ME
Multielement standard in nitric acid 1% M4, M6, Kraft ME
tips + tricks ICP-OES: 259.370 nm is interfered by both, Na and Mg!