sample
type
|
marine
porewater
|
digested
sediment 1g/L
|
groundwater
|
typical
concentration |
1 µmol/L
50 ppb
|
1 mmol/L
50 ppm
|
100 µmol/L
5 ppm
|
Method1 |
photometric
detection |
wavelength |
565
nm |
method used
|
yes
|
no
|
no
|
dilution |
|
|
|
sample size |
1
mL
|
|
|
detction
limit |
20
ppb
|
|
|
standard solutions |
Single
element standard
|
|
|
Method2 |
ICP-OES
|
wavelength |
259.940
nm -> wavelength table |
method used
|
no
|
yes
|
no
|
dilution |
|
-
|
|
sample size |
|
>
2 mL
|
|
detction
limit |
|
20
ppb
|
|
standard solutions |
|
Single
element standard in nitric acid 1%
|
|
Method
3 |
GF-AAS |
wavelength |
248.3
nm |
method used
|
yes
|
yes
|
yes
|
dilution |
1:10
|
-
|
-
|
sample size |
>
100 µL
|
>
100 µL
|
>
100 µL
|
detction
limit |
1
ppb
|
1ppb
|
1
ppb
|
standard solutions |
Single
element standard in nitric acid 1%
|
Single
element standard in nitric acid 1%
|
Single
element standard in nitric acid 1%
|
Method
4 |
ICP-MS |
Mass |
56,
57 -> Mass Table |
method used
|
yes
|
yes
|
yes
|
dilution |
1:100
|
1:10
|
1:10
|
sample size |
>
1 mL
|
>
1 mL
|
>
1mL
|
detction
limit |
10
ppb
|
1
ppb
|
1
ppb
|
standard solutions |
|
|
|
tips
+ tricks |
ICP-OES
234.349nm line is interfered by Ca line |