Uni Bremen Home MARUM - Sediment Geochemistry - Leobener Str - D-28359 Bremen - Germany
Marine Geochemistry - Laboratory Methods

inlab methods main

onboard methods main


Access element methods

Co - Cobalt

sample type

marine porewater
digested sediment 1g/L
groundwater
typical concentration
0.01 µmol/L
1 ppb
0.3 µmol/L
20 ppb

0.01 µmol/L
1 ppb

Method1 ICP-OES
wavelength 228.616 nm -> wavelength table
method used
no
yes
no
dilution
 
-
 
sample size
 
> 2 mL
 
detction limit
 
20 ppb
 
standard solutions
 
Single element standard in nitric acid 1%
 
Method 2 GF-AAS
wavelength 240.7 nm
method used
no
no
yes
dilution    
-
sample size    
500 µL
detction limit    
1 ppb
standard solutions    
Single element standard in nitric acid 1%
Method 3 ICP-MS
Mass 59 -> Mass Table
method used
no
yes
yes
dilution  
1:10
1:10
sample size  
> 1 mL
> 1mL
detction limit  
1 ppb
1 ppb
standard solutions  
Multielement standard in nitric acid 1% M4, M6, Kraft ME
Multielement standard in nitric acid 1% M4, M6, Kraft ME
tips + tricks  

page created by M.Kölling & T.Feseker