Uni Bremen Home MARUM - Sediment Geochemistry - Leobener Str - D-28359 Bremen - Germany
Marine Geochemistry - Laboratory Methods

inlab methods main

onboard methods main


Access element methods

Ca - Calcium

sample type

marine porewater
digested sediment 1g/L
groundwater
typical concentration
10 mmol/L
400 ppm
250 µmol/L
10 ppm

1 mmol/L
40 ppm

Method1 ICP-OES
wavelength 317.932 nm -> wavelength table
method used
yes
yes
yes
dilution
1 : 20 
no
 no
sample size
 > 1 mL
> 5 mL
> 5 mL 
detction limit    
 
standard solutions
single element standard in 1% HNO3
adapted to 1:20 diluted seawater Na-content 
 
 
tips + tricks  

page created by M.Kölling & T.Feseker