Porenwasser
Methoden-Name: Porenwasser 1:10/Ba
Methoden-Beschreibung: PW 1:10 Ba
Element Wellenlänge Name
Ba 493,404 nm Ba 493,408
Ba 455,395 nm Ba 455,403
Resolution: normal Replicates: 3
Read Time (sec.) Auto:      Min. Time: 20
Max. Time: 20
Plasma Parameter: Plas L/min 15
Aux L/min 0,4
Neb L/min 0,82
Power 1200 W
View 11 mm
Pumpen Parameter:
Schlauch: black/black
Sample Flow Rate: 1.2 mL/min
Sample Flush Time (sec.) 12
Sample Flush Rate: 4 mL/min
Read Delay Time (sec.) 60
Wash Rate (mL/min.): 1,2
Wash Time (mL/min.): 70
Process: Background BGC1 BGC2
Correction
Ba 493,408 2-Punkt 0,093 -0,078
Ba 455,403 2-Punkt -0,084 0,084
Peak Algorthms Pts/Peak
Ba 493,408 Area 3
Ba 455,403 Area 3
Calib: Calib Units Calib Conc.
Calib Standard 1 µg/L 50 µg/L
         Standard 2 µg/L 10µg/L
         Standard 3 µg/L 5 µg/L
Calibration and Blank Usage: Blank Action
Element: Ba 493,408 calibrate
Ba 455,403 calibrate
Calibration and Sample Units: Equation  Sample Units: Max.Dec. Max.Signif.
Element: Stellen Figures
Ba 493,408 linear µg/L 3 4
Ba 455,403 linear µg/L 3 4
Options: Auto Integration Data
Replicates
x Means
x Statistics
x Calibrationsummary